Yanislava Karusheva, Matt Ratcliff, Audrey Melvin, Alexander Moerseburg, Naveed Sattar, Peter Barker, Keith Burling, Anna Backmark, Robert Roth, Lutz Jermutus, Esther Guiu-Jurado, Matthias Blueher, Paul Welsh, Marko Hyvönen, Stephen O’Rahilly
MedRxiv preprint DOI: 022.01.03.22268655
Abstract
Genetic variants in proteins can interfere with measurement of their circulating concentrations. Given the growing biomedical importance of GDF-15, we wished to establish whether a common histidine to aspartate variant present in position 6 of the mature GDF-15 protein (H202D variant) interfered with its measurement by two commonly used immunoassays. We first examined the detectability of recombinant monomers, homodimers and heterodimers of GDF-15 by assays and/or reagents used in two widely used immunoassays (Roche Elecsys GDF-15 and the R&D antibody combinations used in their Quantikine and DuoSet ELISAs). The Roche assay detected the H and D containing peptides similarly but the assays based on the R&D reagents consistently underreported concentrations of the D-containing variant peptide. Measurements of plasma concentrations of GDF-15 in genotyped human participants showed that the R&D reagents reported values in heterozygotes were ~25% lower, and in homozygotes, 50% lower than the Roche assay. Continue reading →